FDA Brief

Company record

Intellijoint Surgical, Inc.

0 FDA warning letters since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about Intellijoint Surgical, Inc..

Recalls

ReportedClassFirmProduct and reason
Oct 14 Class II Intellijoint Surgical, Inc.
Device
Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.